ASTM D3078

Gross Leak Tester

ASTM D3078 is a standard test method for determination of gross leaks in flexible pacakging containing headspace gas by bubble emission. The test is conducted by submerging the packages into water within vacuum chamber. Positive pressure inside the package after the vacuum is drawn may force the package to plug small leaks. If bubbles or seepage of fluid within specimen attributable to a leak are observed, then the specimen fails the test.

Apparatus & Materials

1. Vacuum Chamber - a vacuum chamber should be capable of withstanding approximately one atmosphere pressure differential, fitted with a vacuum-tight cover. A vacuum gage, an inlet tube from a source of vacuum, and an outlet tube to the atmosphere sahll be connected to the chamber cover.

2. Immersion Fluids - usally, we use water as the immersion fluid, which does not degrade the package being tested.

Similar Test Method

ASTM F2029 -- Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)

with a customized fixture, Labthink LSSD-01 can conduct the bubble leak test according ASTM F2029

C660B 漏れ&密封強度試験機

C660B 漏れ&密封強度試験機は、食品、医薬品、医療機器、家庭用化学製品、自動車、電子部品、文房具、その他工業製品の梱包の漏れ試験用に設計された試験機です。この機器は、落下および圧縮試験後の試験片の密封性能試験にも使用できます